Welcome to EnviroDIY, a community for do-it-yourself environmental science and monitoring. EnviroDIY is part of WikiWatershed, an initiative of Stroud Water Research Center designed to help people advance knowledge and stewardship of fresh water.
New to EnviroDIY? Start here

Reply To: Negative turbidity readings

Home Forums Mayfly Data Logger Negative turbidity readings Reply To: Negative turbidity readings

#19783
nick
Participant

    Thank you for the information. I will try the voltmeter for confirmation.

    Can a 7-year-old OBS3+ turb sensor go bad if it has been sitting in a climate-controlled room? It was probably used only a couple of times for testing indoors but has been powered off for at least 6 years. I get a positive reading when exposed in the open air, but when I cover it with a towel, I get a negative reading. I’m getting similar results with two different mayfly logger boards, so I think I can rule out a bad analog-to-digital converter.

    Do you have a recommendation for connecting a OBS3+ turb sensor to the current Multipurpose 6-pin Screw Terminal Grove Adapter?

    I have the white cable (high range signal) going to S2, blue (low range) to S1, red (power) to V+. Since there are only 2 grounds on the current grove adapter, how should I plug in the green, black, and clear wires? Should the black (power ground) go into the V+ slot?

    And just to confirm the OBS3+ jumper is set at 5V in the Auxiliary ADC Grove connector while the ClariVue10 jumper is set at 12V in the Digital pin Grove connector.